Prodotti Life Science

Fluke - 9118A Thermocouple Calibration Furnace

Fornace orizzontale ad alta perfoermance per taratura di termocoppie, sino a 1200°C

La fornace 9118A Fluke Calibration è orizzontale con range da 300°C a 1200°C. Impiegata in laboratori per la taratura in comparazione di termocoppie, in differenti settori: aerospaziale, automotive, energia, metallo, plastic.

Le line guida come l’ AMS 2750 e l’ EURAMET cg-8, richiedono che le termocoppie siano tarate nel loro intero range di utilizzo. Il range di lavoro della fornace 9118A da 300°C a 1200°C soddisfa la maggior parte delle applicazioni.

In caso l'inserto standard non dovesse essere utilizzabile, richiedete quotazione per inserto custom.

General specifications

Operating conditions

Operating temperature

5 °C to 40 °C

Storage temperature

–20 °C to 70 °C

Humidity

80 % maximum for temperatures

Altitude

<2,000 m (6,562 feet)

Power requirements

230 V ac (±10 %), 50/60 Hz, 20 A

Heater power

4000 watts at 230 V ac

Over-current protection

System

20 A, 250 V resettable circuit breaker

Main heater fuse

F 12 A, 250 V

Zone heater fuse

F 12 A, 250 V

Computer interface

RS-232 and USB

Display

Monochrome LCD, °C or °F user-selectable

Display resolution

0.1 °C or °F

Size (H x W x L)

400 mm x 337 mm x 700 mm (15.7 in x 13.3 in x 27.6 in)

Net weight

29 kg (63.9 lb) without isothermal block

Isothermal block (optional)

Isothermal block construction

Alumina

Block outer diameter

37 mm (1.5 in)

Block length

380 mm (15.0 in)

Well diameter

(four wells total) 6.7 mm (0.26 in)

Well depth

365 mm (14.4 in)

Net weight

0.84 kg (1.9 lb) isothermal block only

Accuracy specifications

All accuracy specifications except temperature are for a period of 1 year after calibration, at 13 °C to 33 °C

The furnace can be used with or without an Isothermal Block installed. Unique calibration parameters are necessary for each configuration. Calibration does NOT come standard from the factory for each configuration. Ensure that the furnace has been properly calibrated for the desired mode of operation.

Furnace temperature range

300 °C to 1200 °C

Set-point accuracy

± 5 °C

 

Radial uniformity

Temperature

9118A
(14 mm from geometry
central point)

9118A-ITB
(hole to hole)

300 °C

± 0.5 °C

± 0.1 °C

700 °C

± 0.5 °C

± 0.20 °C

1200 °C

± 0.5 °C

± 0.25 °C

Axial uniformity

Temperature

9118A
(± 30 mm axial length from
geometry central point)

9118A-ITB
(at 60 mm from full immersion)

Full range

± 0.25 °C

± 0.2 °C

Temperature stability

Specification

9118A

9118A-ITB

Stability

± 0.2 °C

± 0.1 °C

Stabilization time

2 hours, full range

3 hours at or below 700 °C
2 hours above 700 °C

Note: Temperature stability measured as 2-sigma over 30 minutes

 

9118A

9118A-ITB

Heating time (23  °C to 1200 °C)

40 minutes

45 minutes

Cooling time (1200 °C to 300°C)

180 minutes

200 minutes

 