Strumentazione e Sistemi per Metrologia - Power Supply - T&M - Processo      tel. (+39) 031526566
SPECIFICHE CALIBRATORI PORTATILI CALOG 

CALOG Calibrators - LOOP II Calibrator
    *  Measure, source or simulate 0 to 24 mA
    * Measure 0 to 32 Vdc
    * Continuity test function
    * Increment, step, auto step or ramp with built-in 24V power supply for mA loops
    * Displays values in mA and % or % and mA
    * Logged data is downloaded via the onboard USB port
    * Trend logging with programmable timebase of all measured values
    * Carry case, protective rubber boot, Li-ion battery, charger, test leads and SD card supplied as standard
    * 2 year guarantee against faulty material or workmanship (batteries not included)

   Technical Specifications of the CALOG Calibrators - LOOP II Calibrator
Analogue Input Ranges                   Impedance                                                           Accuracy           Resolution
0 to 24 mA                                       Input Impedance ± 17 ohm, 250mA Fused        0.01%FS         ±1 digit 1µA
0 to 32 V                                         Input Impedance ± 110 kohm                             0.005%FS      ±1 digit  1mV
Continuity with a <100 ohm trigger @ 1 mA. Gives visual and audible confirmation.
      
Analogue Output Ranges                 Maximum Load                                                    Accuracy           Resolution
0 to 24 mA                                        Output load max 500 ohm                                  0.01%FS          ± 1 digit  1µA

 

   CALOG
Calibrators - TEMP Calibrator
   *  Measure, source or simulate 0 to 24 mA
    * Measure and source -10 to 100 mV
    * Measure and source 0 to 400.00 ? and 0 to 2200.0 ?
    * Measure and source types K, J, T, B, R, S, E, N, U and L thermocouples with configurable internal, external or manual
       cold-junction compensation in either °C, °F, Kelvin or Rankine
    * Measure and source Pt50, Pt100, Pt200, Pt500, Pt1000, Ni100 and Ni120 type RTD sensors
    * Continuity test function
    * Increment, step, ramp or valve stroking with built-in 24V power supply for mA loops
    * Datalogging facility via SD card which is downloadable using the USB port
    * Trend logging feature with programmable timebase
    * Carry case, protective rubber boot, Li-ion battery, charger, test leads, CJ Compensation box and
       SD card supplied as   standard
    * 2 year guarantee against faulty material or workmanship (batteries not included)

   Technical Specifications of the CALOG Calibrators - TEMP Calibrator
Analogue Input Ranges                 Impedance                                                   Accuracy                                     Resolution
0 to 24 mA                                 Input Impedance ± 17ohm, 250mAFused      0.02%FS                                  ± 1 digit    1µA
-10 to 100 mV                           Input Impedance > 1mohm                              0.01%FS                                  ± 1 digit    1µV
Thermocouple types are Input Impedance > 1mohm                                         0.01%FS*                       ± 1 digit(excl CJ) 0.1°C
K, J, T, B, R, S, E, N, U and L
0 to 400.00 ohm                                                                                                   0.05%FS                                   ± 1 digit 0.01ohm
0 to 2200.0 ohm                                                                                                   0.05%FS                                   ± 1 digit 0.1ohm
RTD types are Pt50, Pt100, Pt200, Pt500, Pt1000, Ni100 and Ni120 0.05%FS ± 1 digit 0.01°C
Measurement of Ohms and RTD in 2, 3 or 4-wire configuration
Continuity with 100 ohm trigger and 1mA excitation. Gives visual and audible confirmation.
      
Analogue Output Ranges                     Maximum Load                                       Accuracy                                        Resolution
0 to 24 mA                                       Output load max 500 ohm                          0.02%FS                                     ± 1 digit 1µA
-10 to 100 mV                                 Output load min 100 ohm                           0.01%FS                                     ± 1 digit 1µV
Thermocouple types are   0.01%FS* ± 1 digit(excl CJ)0.1°C
K, J, T, B, R, S, E, N, U and L
0 to 400.00 ohm 10 - 400 ohm                                                                            0.05%FS                                    ± 1 digit 0.01ohm
0 to 2200.0 ohm 10 - 2k2 ohm                                                                            0.05%FS                                    ± 1 digit 0.1ohm
RTD types are Pt50, Pt100, Pt200, Pt500, Pt1000, Ni100 and Ni120 0.05%FS ± 1 digit 0.01°C



   CALOG
Calibrators - Process Calibrator
    *  Measure, source or simulate 0 to 24 mA
    * Measure 0 to 32 Vdc and Source 0 to 12 Vdc
    * Measure and source -10 to 100 mV
    * Continuity test function
    * Measure and source frequency from 0 to 20kHz
    * Measure and source counts up to 8 digits (99 999 999)
    * Increment, step, auto step or ramp with built-in 24V power supply for mA loops
    * Displays values in mA and % or % and mA
    * Simultaneously measure and source all mA, mV and V combinations (not isolated)
    * This unit covers most requirements of PLC and SCADA applications.
    * Carry case, protective rubber boot, rechargeable battery, charger and test leads supplied as standard
    * 2 year guarantee against faulty material or workmanship (batteries not included)
   Technical Specifications of the CALOG Calibrators - Process Calibrator
Analogue Input Ranges                                   Impedance                                                       Accuracy                          Resolution
0 to 24 mA                                                  Input Impedance ± 17 ohm, 250mA Fused       0.01% FS                     ± 1 digit 1µA
0 to 32 V                                                     Input Impedance ±110 k
ohm                            0.005% FS                   ± 1 digit 1mV
-10 to 100 mV                                            Input Impedance >1 M
ohm                                 0.005% FS                   ± 1 digit1µV
0.00 to 100.00 Hz, trigger level up to 12.0 V Input Impedance 110 k
ohm                         0.01% FS                   ± 1 digit 0.01Hz
Pulses / min max = 6000.00 and Pulses / hour max = 360 000.00
0 to 20 000 Hz, trigger level up to 12.0 V Input Impedance 110 k
ohm 0.01% FS ± 1 digit 1Hz
Pulses / Min max = 1 200 000 and Pulses / Hour max = 72 000 000
0 to 99 999 999 counts, trigger level up to 12.0 V in 0.1 V steps
Continuity with a <100
ohm trigger using 1 V @ 1 mA. Gives visual and audible confirmation.
      
Analogue Output Ranges                       Maximum Load                                                           Accuracy                         Resolution
0 to 24 mA                                             Output load max 500
ohm                                         0.01% FS                      ± 1 digit 1µA
0 to 12 V                                                Output load min 500
ohm                                          0.01% FS                      ± 1 digit 1mV
-10 to 100 mV                                       Output load min 100
ohm                                         0.01% FS                      ± 1 digit 1µV
0.50 to 100.00 Hz, trigger level up to 12.0 V 0.01% FS ± 1 digit 0.01Hz
1 to 20 000 Hz, trigger level up to 12.0 V   0.01% FS ± 1 digit 1Hz